Software Update (Rev 2.2.804.158)

Announcing Software Revision 2.2.804.158 for EVOS M7000 Imaging System

Revision 2.2.804.158 provides support for EVOS second-generation onstage incubator (OSI-2) as well as improved functionality for the original onstage incubator (OSI).

Please review the Installation Instructions below before updating your software.

New features

OSI-2

  • Improved control over onstage incubator environmental settings including temperature, humidity, and the new capability to select from an external air source or use the OSI-2’s internal air pump.
  • Improved layout of real-time incubator readings during a time-lapse experiment.
  • Prompt for selecting gas lines when configuring the OSI-2.

OSI

  • Improved layout of real-time incubator readings during a time-lapse experiment.
  • Prompt for selecting gas lines when configuring the OSI.

Installation Instructions

Software Update (Rev 2.2.804.158)

IMPORTANT! Restart your system prior to updating to optimize the update process.

Update from USB

Uninstall the Current Software

  1. Go to Windows Start and select Settings, then click on Apps
  2. Under Apps and Features select M7000 from the list of installed applications
  3. Click “Uninstall” to uninstall the software
    Download and Install the Software Update
  4. Insert a fresh USB drive with at least 240 MB available space.
  5. Right-click the “Download” link for Software Update.
  6. Select “Save Link as”.
  7. Select the USB Drive and press Save.
  8. Plug the USB flash drive into the M7000 computer
  9. Locate the update file from your USB drive in Windows Explorer and copy it to the hard drive.
  10. Double-click the update file from the hard drive to install the update.
  11. A WindowsTM dialog box will pop up. Click Next to proceed with the software update.
  12. The screen will display the update progress. When the update is complete, click OK.
  13. Shut down the computer (Windows Start -> Power icon -> Shut down).
  14. Power up the computer and launch the M7000 software.
    Recalibrate Your Instrument
  15. Open the Settings Tab.
  16. Click on Objectives, recalibrate your objectives, then click Done at the bottom.
  17. Click on Stage Calibration, recalibrate the stage, then click Done at the bottom.
  18. Recalibrate any vessels by mounting them on the stage, click Vessels, select Calibrate Vessel, follow the instructions, and when finished click Done at the bottom.

IMPORTANT! If you experience any issues with the installation or find that it is not completing as expected, please reboot your instrument and contact Technical Support.

The number for your local Technical Support Team can be found by navigating to our home page.